04-Inner-Installation

Design Qualification

  • Defines the functional and operational requirement and specification of the equipment and details the conscious decisions in the selection of the supplier.

Installation Qualification

  • Confirms that all the equipment was installed to specifications. All of the model and serial numbers are documented.

Operational Qualification

Verification of operator control feature, safety features, software feature, Stan by feature with alarms action.

Performance Qualification

Confirms that the equipment working within operating range at stipulated product load.

Pre Installation Requirement

Surrounding Area Clean & free from dust
Temp 20 ℃ ~ 35 ℃ (Recommended using AC)
Level Horizontally leveled
Clearance Space Right & left Side Minimum 500mm
Back Side Minimum 500mm
Top Side Minimum 500mm
Door Open Area Greater than doors width size of equipment
Refrigeration unit (for Walk in type chamber) Distance between equipment and compressor unit should not exceed 45 feet
3“ diameter hole required at wall for passing copper pipe of split type refrigeration unit.
Proper shade is required to protect compressor unit from rain water.
Required stabilized power supply for equipment For non walk-in type Single phase, 230+10 VAC, 50 Hz, 20 Amp with proper earthing
For walk-in type Three phase, 440+10 VAC, 50 Hz, 32 Amp with proper earthing
Required UPS for PLC & Control system Single phase, 230+10 VAC, 50 Hz, 250 VA
Required LAN port [I/O] for communication of equipment and software Information outlet [I/O] for each equipment and distance between equipment & port should not exceed 2.5 meters.
Drainage Level Concealed (below ground level)
Continuous water supply (For Temperature & Humidity base Equipment) Pipe specification 3“ diameter PVC
Water quality Dust free Demineralized or soft water
Pressure 1 kg/cm ~ 1 Bar ~ 15 psi approx
Valve diameter ½ “BSP