Vertical Laminar Air Flow Use
Kesar Control Systems

Vertical Laminar Air Flow

Introduction

Kesar Control Systems manufactures high-quality Vertical Laminar Air Flow (VLF) units designed to provide a sterile and particle-free working environment for sensitive laboratory and pharmaceutical applications. The system delivers HEPA-filtered air vertically from top to bottom, creating a unidirectional airflow that protects products from airborne contamination. This controlled airflow sweeps particles away from the work surface and maintains ISO Class 5 clean air conditions.

Construction & Design

  • Cabinet fabricated in SS-304 / SS-316.
  • Smooth inner surfaces to prevent contamination.
  • Work table made of stainless steel perforated sheet.
  • Transparent side panels for clear visibility.
  • Designed according to GMP guidelines.
  • Corrosion-resistant and easy to disinfect.

Filtration System

Primary Filtration:

Non-woven Pre-Filter (EU-04 / G4).

Efficiency: 95% down to 5 μm.

Secondary Filtration:

Mini-Pleat HEPA Filter – H14 Grade

Efficiency: 99.997% for particles ≥0.3 μm

Control & Instrumentation

Feather-touch LED controller panel

Magnehelic differential pressure gauge

Digital hour meter

Blower ON/OFF control

UV lamp timer control

LED lighting control

Optional filter blockage alarm

Lighting System

Fluorescent / LED light with minimum 300–1000 lux illumination

UV Germicidal Lamp for sterilization during idle periods

Quick Contact

Get Solution

Contact Kesar Control today to learn more about their Products and get a quote.

Contact Now

Features

  • ISO Class 5 cleanliness according to ISO 14644-1.
  • Vertical unidirectional laminar airflow.
  • Mini-Pleat HEPA Filter H-14 with 99.997% efficiency at 0.3 μm.
  • Two-stage filtration system (Pre-filter + HEPA filter).
  • Stainless Steel SS-304 / SS-316 construction.
  • Touch LED controller for blower, lighting and pressure display.
  • Magnehelic differential pressure gauge.
  • UV Germicidal lamp for sterilization.
  • LED / Fluorescent lighting.
  • Low noise operation ≤ 65 dB.
  • GMP compliant design.
  • Easy cleaning and maintenance.

Applications

  • Pharmaceutical laboratories
  • Microbiology laboratories
  • Biotechnology research
  • Tissue culture work
  • Sterile drug preparation
  • Electronics assembly
  • Food quality testing

Technical Specification

Air Cleanliness ISO Class 5
Airflow Direction Vertical
Air Velocity 0.45 ± 0.05 m/s
HEPA Filter Efficiency 99.997% at 0.3 μm
Pre-Filter Efficiency 95% at 5 μm
Noise Level ≤ 65 Db
Lighting ≥ 300 Lux
Power Supply Single Phase 220–230 V, 50 Hz

Explore More

Similar Products

Request A Quote